Skip to airway content Airway of the MonthReview the case. Understand the airway. Build the skill.

EtCO₂: tube problem

Troubleshooting review

EtCO₂: tube problem

When waveform is absent, do not trust the airway until proven.

Check

  • If an immediate equipment correction does not restore sustained exhaled CO₂, remove the tube and ventilate with 100% oxygen by mask or SGA. Retain it only with a reasonable basis for tracheal placement and serious danger from removal.
  • If retained, promptly exclude esophageal placement using repeat laryngoscopy, flexible bronchoscopy or trained airway ultrasound. Remove the tube if placement remains uncertain, sustained CO₂ cannot be restored or oxygenation worsens.
  • Chest rise, auscultation and chest X-ray cannot exclude esophageal intubation.
  • Once esophageal placement is excluded, reassess tube depth, obstruction, cuff and circuit.

Clinical review / use limits

Authored by: Airway of the Month education team. Last reviewed: July 2026. Audience: EM learners, faculty, EMS, RT, simulation educators, and airway teams.

About this resource
Website creator: Kyle Brown, MD Airway educational contributors: Andrew Pirotte, MD; Kyle Brown, MD; Reba Hodge, MD; Amanda Vanderwerf, PharmD, BCEMP; Morgan Kimball, PharmD, BCEMP; Joshua Mohess Clinical review: Emergency medicine faculty review completed. Medication content reviewed by Amanda Vanderwerf, PharmD, BCEMP, and Morgan Kimball, PharmD, BCEMP. Full-site review: 2026-07-23 Medication review update: 2026-11-06 Next full-site review: 2027-01-23 Designed for: EM residents, emergency physicians, EMS clinicians, airway educators, and simulation faculty Important: Educational resource and clinical-reference guide only; not a bedside order set or substitute for local protocol, medical direction, or clinical judgment. Website contact: [email protected]